1. Safety of electromagnetic articulography in patients with pacemakers and implantable cardioverter-defibrillators: research note.
- Author
-
Joglar JA, Nguyen C, Garst DM, and Katz WF
- Abstract
PURPOSE: Electromagnetic articulography (EMA) uses a helmet to create alternating magnetic fields for tracking speech articulator movement. An important safety consideration is whether EMA magnetic fields interfere with the operation of speakers' pacemakers or implantable cardioverter-defibrillators (ICDs). In this investigation, individuals with pacemaker/ICD devices were exposed to EMA fields under controlled conditions while potential interference was examined. METHOD: Twelve adults with pacemaker/ICD devices from 3 major manufacturers were assessed for device function before, during, and after exposure to magnetic fields from a Carstens AG100 EMA system. Potential interference was probed, with EMA transmitters positioned at varying distances from the implantable devices and with the EMA system set at different operating strengths. RESULTS: No adverse affects in device operation were observed under any conditions. The only potential complication was temporary telemetry-link interference during device testing in some cases. CONCLUSION: The results suggest that EMA technology may be safely used with patients who have pacemakers and ICDs. However, the present findings do not rule out potential interference with other pacemaker/ICD manufacturers or with different articulography systems. Precautions are suggested for testing individuals with pacemaker/ICDs under EMA conditions. [ABSTRACT FROM AUTHOR]
- Published
- 2009
- Full Text
- View/download PDF