1. Accuracy of a Factory-Calibrated Continuous Glucose Monitor in Individuals With Diabetes on Hemodialysis
- Author
-
Orianne Villard, Marc D. Breton, Swati Rao, Mary K. Voelmle, Morgan R. Fuller, Helen E. Myers, Ryan K. McFadden, Zander S. Luke, Christian A. Wakeman, Mary Clancy-Oliveri, Ananda Basu, and Meaghan M. Stumpf
- Subjects
Blood Glucose ,Advanced and Specialized Nursing ,Diabetes Mellitus, Type 1 ,endocrine system diseases ,Renal Dialysis ,Blood Glucose Self-Monitoring ,Endocrinology, Diabetes and Metabolism ,Internal Medicine ,Humans ,Reproducibility of Results ,nutritional and metabolic diseases - Abstract
OBJECTIVE Continuous glucose monitoring (CGM) improves diabetes management, but its reliability in individuals on hemodialysis is poorly understood and potentially affected by interstitial and intravascular volume variations. RESEARCH DESIGN AND METHODS We assessed the accuracy of a factory-calibrated CGM by using venous blood glucose measurements (vBGM) during hemodialysis sessions and self-monitoring blood glucose (SMBG) at home. RESULTS Twenty participants completed the protocol. The mean absolute relative difference of the CGM was 13.8% and 14.4%, when calculated on SMBG (n = 684) and on vBGM (n = 624), and 98.7% and 100% of values in the Parkes error grid A/B zones, respectively. Throughout 181 days of CGM monitoring, the median time in range (70–180 mg/dL) was 38.5% (interquartile range 29.3–57.9), with 28.7% (7.8–40.6) of the time >250 mg/dL. CONCLUSIONS The overall performance of a factory-calibrated CGM appears reasonably accurate and clinically relevant for use in practice by individuals on hemodialysis and health professionals to improve diabetes management.
- Published
- 2022
- Full Text
- View/download PDF