1. Occlusion following the deployment of MANTA VCD after TAVR.
- Author
-
Hassan MF, Mixon T, Gray WT, and Reddy R
- Subjects
- Arterial Occlusive Diseases diagnostic imaging, Humans, Middle Aged, Postoperative Complications diagnostic imaging, Severity of Illness Index, Aortic Valve Stenosis surgery, Arterial Occlusive Diseases etiology, Femoral Artery, Postoperative Complications etiology, Transcatheter Aortic Valve Replacement, Vascular Closure Devices adverse effects
- Abstract
Background: We present a complication following the deployment of the MANTA vascular closure device (VCD) following a transcatheter aortic valve replacement procedure which resulted in occlusion of the common femoral artery., Aims: To address possible complications associated with MANTA VCD., Results: This was addressed by ballooning the site from the contralateral side which re-established flow., Conclusion: We believe this is the first report to address this kind of complication and may prove useful as more of the MANTA devices are being used in multiple procedures., (© 2020 Wiley Periodicals LLC.)
- Published
- 2020