1. Integration of beta counting system with ORION-DSP coupled to NaI(Tl) detector—a comparative study with an HPGe detector.
- Author
-
Rupa, Nyapathy, Kumar, Pavan, Shankar, Madhavi, Reddy, Ram Mohan, Rao, Bala Bhaskara, and Mahalakshmi, Balasubramanian
- Subjects
GAMMA ray detectors ,GERMANIUM radiation detectors ,DETECTORS ,COUNTING ,COMPARATIVE studies - Abstract
The integration of the ORION digital signal processing-based MCA system coupled with a 3″ × 3″ NaI(Tl) detector assembly with a GM detector for counting beta (β) has eliminated the need for a standalone β-γ method in which U
3 O8 is determined by gross β and gross γ counting. Uraniferous and mixed U-Th samples were taken up for study and compared with the results obtained from Canberra p-type coaxial high resolution gamma ray spectrometry detector. In uraniferous samples, U3 O8 values obtained are within ±10%, whereas in the case of mixed U-Th samples, U3 O8 values are within ±15%. Regression graphs drawn between the outcomes from the two analytical systems indicate R2 > 0.95 for Ra(eU3 O8 ) and ThO2 . In uraniferous samples, the R2 value for U3 O8 was found to be > 0.99, but in mixed U-Th samples, it is 0.92. The closeness of agreement between the results obtained from two methods at various concentrations over the analytical range shows that the integrated system is suitable for the quantitative determination of eU3 O8 , U3 O8 , Ra(eU3 O8 ), ThO2 and K in geological rock samples. [ABSTRACT FROM AUTHOR]- Published
- 2024
- Full Text
- View/download PDF