1. Clinical Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in Kenya.
- Author
-
Francis Angira, Benta Akoth, Paul Omolo, Valarie Opollo, Scott Bornheimer, Kevin Judge, Henok Tilahun, Beverly Lu, Imelda Omana-Zapata, and Clement Zeh
- Subjects
Medicine ,Science - Abstract
The BD FACSPresto™ Near-Patient CD4 Counter was developed to expand HIV/AIDS management in resource-limited settings. It measures absolute CD4 counts (AbsCD4), percent CD4 (%CD4), and hemoglobin (Hb) from a single drop of capillary or venous blood in approximately 23 minutes, with throughput of 10 samples per hour. We assessed the performance of the BD FACSPresto system, evaluating accuracy, stability, linearity, precision, and reference intervals using capillary and venous blood at KEMRI/CDC HIV-research laboratory, Kisumu, Kenya, and precision and linearity at BD Biosciences, California, USA.For accuracy, venous samples were tested using the BD FACSCalibur™ instrument with BD Tritest™ CD3/CD4/CD45 reagent, BD Trucount™ tubes, and BD Multiset™ software for AbsCD4 and %CD4, and the Sysmex™ KX-21N for Hb. Stability studies evaluated duration of staining (18-120-minute incubation), and effects of venous blood storage
- Published
- 2016
- Full Text
- View/download PDF