1. Preliminary validation system for cuffless blood pressure measurement
- Author
-
Gonçalves Seabra Ana Carolina, Ferreira da Silva Alexandre, Stieglitz Thomas, and Belen Amado-Rey Ana
- Subjects
bp monitoring ,cuffless ,mimicked arm ,stiffness ,ex-vivo validation ,Medicine - Abstract
The development of a non-invasive, cuffless, continuous, wearable device for the measurement of blood pressure is a complex endeavour due to the high specificity at each measuring site and the need for high accuracy. Proof-ofconcept and validation of a prototype should be performed at an early stage for functionality assessment. Additionally, the emergence of biological computer models allows for in-silico research, which results should be verified in a practical experiment. To grant an optimal preliminary assessment of a prototype, this work aimed to develop and validate accurate in-vitro and ex-vivo arterial models, with simple construction and easily available components. The comparison between a silicone tube and a porcine artery as a mimicked human radial artery was based on the stiffness parameter. Flow pressure is controlled by a centrifugal heart-like pump. Pressure values are extracted with ultrasound and a commercial piezoresistive pressure sensor is used for pressure validation. The porcine artery showed much more realistic stiffness values (·=15.360) than the silicon tube (·=543.420), which was very stiff in comparison to the typical in-vivo radial artery stiffness (·=9.5). The decrease in stiffness of 97.173 % (from the silicone tube to the porcine artery) led to an acute decrease in the derived pressure error. This work serves as guidelines for the development of a low-budget arm phantom, as the simple setup allowed for a primary validation of a proof-of-concept ultrasound-based sensor for the measurement of pressure.
- Published
- 2022
- Full Text
- View/download PDF