1. Falsely elevated plasma ACTH levels measured by the Elecsys assay related to heterophilic antibody in a case of secondary adrenocortical insufficiency
- Author
-
Kazuhisa Tsukamoto, Yuta Okawa, Masumi Ogawa, Koji Morita, Michiko Kimura, Yutaka Oki, Toshio Ishikawa, and Naoki Edo
- Subjects
endocrine system ,medicine.medical_specialty ,Prednisolone ,Endocrinology, Diabetes and Metabolism ,030209 endocrinology & metabolism ,Glomerulonephritis, Membranous ,Primary Adrenal Insufficiency ,03 medical and health sciences ,0302 clinical medicine ,Endocrinology ,Adrenocorticotropic Hormone ,Membranous nephropathy ,Internal medicine ,Secondary adrenocortical insufficiency ,Adrenal insufficiency ,Humans ,Medicine ,Elevated plasma ACTH ,biology ,medicine.diagnostic_test ,business.industry ,ACTH stimulation test ,Middle Aged ,medicine.disease ,030220 oncology & carcinogenesis ,Polyethylene glycol precipitation ,biology.protein ,Biological Assay ,Female ,Antibody ,business ,hormones, hormone substitutes, and hormone antagonists ,Adrenal Insufficiency - Abstract
A 49-year-old woman with membranous nephropathy was referred to our hospital during the tapering of oral prednisolone, because of suspicion of primary adrenal insufficiency based on a plasma ACTH level of 399.1 pg/mL in the Elecsys assay and a serum cortisol level of 3.1 μg/dL. A rapid ACTH stimulation test revealed a suboptimal response, whereas a prolonged ACTH simulation test showed a sufficient increase in her urinary free cortisol. Also, big ACTH was not detected by gel exclusion chromatography. Therefore, we speculated that ACTH levels were falsely elevated due to some interference substances. Pretreatment of her plasma with either polyethylene glycol precipitation or a heterophilic blocking tube substantially reduced her ACTH values. When either the Immulite ACTH II or the TOSOH II ACTH was tried instead of the Elecsys ACTH, her plasma ACTH values turned out to be lower and appropriate for her clinical status. These results indicated that heterophilic antibodies interfered only with the Elecsys ACTH assay presumably by bridging the capture and tracer antibodies. To our knowledge, this is the first case in which the Elecsys ACTH assay yielded falsely elevated results. Regardless of the measurement system used, if there is a discordance between assay results and clinical findings, it should be considered to adopt additional procedures and/or another assay.
- Published
- 2019
- Full Text
- View/download PDF