1. Development of Rb fountain clock for time keeping
- Author
-
Chen Weiliang, Fang Fang, Liu Kun, Zheng Fasong, Dai Shaoyang, Zuo Yani, and Li Tianchu
- Subjects
fountain clock ,time keeping ,Rb ,double metal cavity ,instability ,Physics ,QC1-999 - Abstract
A new Rb fountain clock was built at the National Institute of Metrology with the aim to realize robustness and excellent long-term instability for time keeping. While some basic designs have been adopted from our Cs fountain clocks, new features have been included for improved performance. A double-metal interrogation microwave cavity with a thermal expansion self-compensating mechanism was used to reduce the clock sensitivity to temperature fluctuations for the first time, to our knowledge. Moreover, a compact optical system was developed to ensure robustness. These features dramatically increased the ambient temperature range. The developed Rb fountain clock achieved a typical fractional frequency instability of 1.49×10−13(τ/s)−1/2within 10 days of continuous operation without any intervention, demonstrating the potential for time keeping.
- Published
- 2022
- Full Text
- View/download PDF