1. Comparison of laser power output from the fiber tip during endovenous laser ablation against displayed power and the "first treatment" effect.
- Author
-
Davies CE, Riley MI, Dabbs EB, and Whiteley MS
- Subjects
- Humans, Lasers, Semiconductor therapeutic use, Models, Biological, Varicose Veins physiopathology, Endovascular Procedures instrumentation, Endovascular Procedures methods, Laser Therapy instrumentation, Laser Therapy methods, Optical Fibers, Varicose Veins therapy
- Abstract
Objective: International guidelines recommend endovenous laser ablation as one of the first-line treatments for truncal venous reflux associated with varicose veins. Clinicians use linear endovenous energy density to measure the energy used during treatment. The aim of this study was to investigate the power output from the fiber tip and to see if this changed with use., Methods: We placed 15 mL of water in a thermally insulated 25-mL beaker. Two thermocouples were placed equidistantly adjacent to the laser fiber tip. A 1470-nm laser was fired at 5W for 2 minutes and the temperature change was measured. Three fibers were used on different days to allow the laser to cool for 24 hours between fibers. Each fiber was tested three times in a row. We also tested the "first treatment" effect by comparing the power output when the fiber was fired immediately after the laser was switched on, compared with treatments when the laser had been switched on for 1 hour. To assess whether this was due to the console being "cold" on the first firing of the day, we repeated the experiment having switched on the laser console 1 hour before firing to "prewarm" the console. However, the diode was not fired during this hour. To measure fiber degradation, three runs of the experiment were performed successively before firing the laser continuously for 20 minutes, then three more runs were conducted, resulting in delivery of approximately 10,000 J., Results: The actual power output seemed to be lower than the console suggested. The power output from the first fiber used in a succession of three with the same laser had a significantly lower power output than the following two runs (P = .0004 and P < .0001, respectively). When the laser was prewarmed for 1 hour without firing, no change in this output pattern was noted (P = .293). Fiber degradation was not found in any of the fibers that were tested within the maximum recommended for the fiber (10,000 J)., Conclusions: The first use of a fiber in a treatment session has a significantly lower power output from the treatment tip than subsequent uses, even if the machine is prewarmed and the console displays the same power for each. The authors believe that this phenomenon is due to the diode being less efficient when first used after switching it on. Clinicians need to be aware that the true power output at the fiber tip may not be as indicated by the console display and may be variable during a treatment session., (Copyright © 2020 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.)
- Published
- 2021
- Full Text
- View/download PDF