1. Long-term telemetric intracranial pressure monitoring for diagnosis and therapy optimisation of idiopathic intracranial hypertension
- Author
-
Victor F. Velazquez Sanchez, Giath Al Dayri, and Christoph A. Tschan
- Subjects
Idiopathic intracranial hypertension ,Intracranial pressure monitoring ,Telemetric intracranial pressure monitoring ,Home-monitoring of ICP ,Home-Telemonitoring of ICP ,Neurology. Diseases of the nervous system ,RC346-429 - Abstract
Abstract Background Idiopathic intracranial hypertension (IIH) is a disease which is difficult to diagnose and moreover difficult to treat. We developed a strategy for long-term telemonitoring of intracranial pressure (ICP), by incorporation of the NEUROVENT®-P-tel System, with the goal of improved diagnosis and consequent therapy of this disease. We highlight the results obtained through this approach. Methods Twenty patients with suspected IIH who were treated in our hospital from August 2014 to October 2020 (16 females, 4 males, median age 36,6 years), were assigned to one of two ICP monitoring settings, “Home-Telemonitoring” (n = 12) and “Home-Monitoring” (n = 8). The ICP data were analysed and used conjointly with the accompanying clinical picture for establishment of IIH diagnosis, and telemonitoring was resumed for therapy optimisation of confirmed cases. Results The diagnosis of IIH was confirmed in 18 of the 20 patients. Various surgical/interventional treatments were applied to the confirmed cases, including ventriculoperitoneal (VP) shunting (n = 15), stenting of the transvers venous sinus (n = 1), endoscopic third ventriculostomy (ETV) (n = 1), and ETV in combination with endoscopic laser-based coagulation of the choroid Plexus (n = 1). Optimal adjustment of the implanted shunt valves was achieved with an average valve opening pressure of 6,3 ± 2,17 cm H2O for differential valves, and of 29,8 ± 3,94 cm H2O for gravitational valves. The Home-Telemonitoring setting reduced consequent outpatient visits, compared to the Home-Monitoring setting, with an average of 3,1 visits and 4,3 visits, respectively. No complications were associated with the surgical implantation of the P-tel catheter. Conclusion This study offers insight into the use of long-term ICP monitoring for management of IIH patients in combination with dual-valve VP shunts. The use of NEUROVENT® P-tel system and potentially other similar fully implantable ICP-monitoring devices, albeit invasive, may be justified in this complex disease. The data suggest recommending an initial adjustment of dual-valve VP-shunts of 30 and 6 cm H2O, for gravitational and differential valves, respectively. Further research is warranted to explore potential integration of this concept in IIH management guidelines.
- Published
- 2021
- Full Text
- View/download PDF