1. Enzymatische Bestimmung von Xylit und Sorbit nach dem Endwertverfahren
- Author
-
Wagner K, Bässler Kh, and Schönerstedt B
- Subjects
chemistry.chemical_classification ,Semicarbazide ,Chromatography ,Coefficient of variation ,education ,Biochemistry (medical) ,Clinical Biochemistry ,Dehydrogenase ,General Medicine ,Xylitol ,Lower limit ,chemistry.chemical_compound ,Enzyme ,chemistry ,Endpoint Determination ,Sorbitol - Abstract
An enzymatic endpoint determination of xylitol and sorbitol by the use of L-iditol dehydrogenase is described. Despite the use of pH 9 and the addition of semicarbazide, the reaction goes only 80% to completion. Therefore a standard value has to be used. The lower limit for measurement of concentrations in serum is around 0.1 mmol/l. The most favorable range is between 0.4 and 1.2 mmol/l with a coefficient of variation of below 4%. Sensitivity and precision is higher than with the kinetic method described earlier. Further simplification for the routine control of infusion therapy with polyols is suggested.
- Published
- 1978