101. Comparison of pediatric lens scattered dose measurements between axial 40-mm and helical 160-mm detector width computed tomography scan modes.
- Author
-
Masuda T, Kiguchi M, Fujioka C, Oku T, Ishibashi T, Katsunuma Y, Yoshitake T, Abe S, and Awai K
- Subjects
- Humans, Male, Female, Tomography, Spiral Computed methods, Phantoms, Imaging, Radiometry, Radiation Exposure, Infant, Newborn, Infant, Child, Preschool, Lens, Crystalline diagnostic imaging, Lens, Crystalline radiation effects, Scattering, Radiation, Tomography, X-Ray Computed methods, Radiation Dosage
- Abstract
Background: Reports comparing field lens doses between helical scans with a 40-mm detector width and axial scans with a 160-mm detector width using different computed tomography (CT) scanners are currently scarce., Objective: To compare scatter doses for lenses between a helical scan with a 40-mm detector width and an axial scan with a 160-mm detector width when using different CT scanners in the context of pediatric chest examinations., Materials and Methods: Two different CT machines were used: Revolution CT (GE Healthcare, Waukesha, WI) with a 256-row, 0.625-mm multidetector; and Aquilion ONE GENESIS Edition (Canon Medical Systems, Otawara, Japan) with a 320-row, 0.5-mm multidetector. Three pediatric anthropomorphic phantoms were used, with optically stimulated luminescence dosimeters (OSLDs) placed on the left and right lenses. The scatter dose values measured by the OSLDs were compared between a helical scan with a 40-mm detector width and an axial scan with a 160-mm detector width during pediatric chest CT examinations., Results: Median equivalent doses for the helical and axial scans were 0.12 and 0.12 mSv/mGy for the newborn, 0.17 and 0.16 mSv/mGy for the 1-year-old, and 0.18 and 0.15 mSv/mGy for the 5-year-old, respectively, when using the Revolution CT. With the Revolution CT, no significant differences were observed in the scatter doses between helical and axial scans in the newborn and 1-year-old phantoms. However, the lens scatter dose for the helical scan was approximately 20-35% higher than that for the axial scan in the 5-year-old phantom (P<0.01). The median equivalent doses of eye lenses for the helical and axial scans were 0.12 and 0.07 mSv/mGy for the newborn, 0.07 and 0.05 mSv/mGy for the 1-year-old, and 0.14 and 0.12 mSv/mGy for the 5-year-old, respectively, when using the Aquilion ONE. With the Aquilion ONE, lens scatter doses for the helical scan were approximately 70%, 40%, and 30% higher in the newborn, 1-year-old, and 5-year-old phantoms, respectively, than those for the axial scan (P<0.01)., Conclusions: When using the Aquilion ONE, lens scatter doses for the helical scan were significantly higher in all three phantoms than those for the axial scan. In contrast, when using the Revolution CT, the lens scatter dose for the helical scan was significantly higher in the 5-year-old phantom than that for the axial scan. These results suggest that although scattered doses may vary with respect to the CT scanner and body size, they are generally lower in the case of axial scans., (© 2024. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.)
- Published
- 2024
- Full Text
- View/download PDF