51. Rate and nature of complications with leadless transcatheter pacemakers compared with transvenous pacemakers: results from an Italian multicentre large population analysis.
- Author
-
Palmisano P, Facchin D, Ziacchi M, Nigro G, Nicosia A, Bongiorni MG, Tomasi L, Rossi A, De Filippo P, Sgarito G, Verlato R, Di Silvestro M, and Iacopino S
- Subjects
- Humans, Treatment Outcome, Equipment Design, Time Factors, Arrhythmias, Cardiac diagnosis, Arrhythmias, Cardiac therapy, Arrhythmias, Cardiac etiology, Pacemaker, Artificial adverse effects
- Abstract
Aims: The safety and efficacy of leadless intracardiac-permanent pacemaker (L-PM) have been demonstrated in multiple clinical trials, but data on comparisons with conventional transvenous-permanent pacemaker (T-PM) collected in a consecutive, prospective fashion are limited. The aim of this analysis was to compare the rate and the nature of device-related complications between patients undergoing L-PM vs. T-PM implantation., Methods and Results: Prospective, multicentre, observational project enrolling consecutive patients who underwent L-PM or T-PM implantation. The rate and nature of device-related complications were analysed and compared between the two groups. Individual 1:1 propensity matching of baseline characteristics was performed. A total of 2669 (n = 665 L-PM) patients were included and followed for a median of 39 months, L-PM patients were on average older and had more co-morbidities. The risk of device-related complications at 12 months was significantly lower in the L-PM group (0.5% vs. 1.9%, P = 0.009). Propensity matching yielded 442 matched pairs. In the matched cohort, L-PM patients trended toward having a lower risk of overall device-related complications (P = 0.129), had a similar risk of early complications (≤30 days) (P = 1.000), and had a significantly lower risk of late complications (>30 days) (P = 0.031). All complications observed in L-PM group were early. Most (75.0%) of complications observed in T-PM group were lead- or pocket-related., Conclusion: In this analysis, the risk of device-related complications associated with L-PM implantation tended to be lower than that of T-PM. Specifically, the risk of early complications was similar in two types of PMs, while the risk of late complications was significantly lower for L-PM than T-PM., Competing Interests: Conflict of interest: P.P. reports no conflicts with regard to the content of the present work, outside this work small speaker fees from Bayer, Boehringer Ingelheim, Bristol-Myers Squibb, and Novartis; M.Z. reports no conflicts with regard to the content of the present work, outside this work small speaker fees from Abbott, Boston Scientific, and Biotronik. No other conflicts of interest were reported., (© The Author(s) 2022. Published by Oxford University Press on behalf of the European Society of Cardiology. All rights reserved. For permissions, please email: journals.permissions@oup.com.)
- Published
- 2023
- Full Text
- View/download PDF