1. Dosimetry evaluation of 192Ir Flexisource using gafchromic EBT3 film.
- Author
-
Oustous, Aziz, Airouss, Zouhair, EL Attaoui, Yacine, Oustous, Yassine, Khalis, Mohammed, and Sebihi, Rajaa
- Subjects
MEDICAL protocols ,RADIOTHERAPY ,PARTICLE accelerators ,RADIATION dosimetry ,RADIOISOTOPE brachytherapy ,MEDICAL societies ,DESCRIPTIVE statistics ,SIMULATION methods in education ,IMAGING phantoms ,TUMORS ,RADIATION doses ,MEDICAL equipment reliability ,CALIBRATION ,SENSITIVITY & specificity (Statistics) - Abstract
Introduction: Brachytherapy is a widely used cancer treatment technique requiring precise dose delivery for optimal outcomes. This study evaluates the dosimetric properties of the ${{\rm{\;}}^{192}}$ Ir Flexisource using Gafchromic EBT3 film, adhering to American Association of Physicists in Medicine guidelines. Methods: The EBT3 film was calibrated with a linear accelerator to establish the relationship between optical density and absorbed dose. Measurements were conducted in a water-equivalent solid phantom to simulate human tissue. The dosimetric properties measured included the dose rate constant, radial dose function and anisotropy function. Results: The results showed a close alignment with reference data: a maximum deviation of 3·1% for the dose rate constant, 2·3% for the radial dose function and 5·67% for the anisotropy function. These deviations are within acceptable limits, indicating high accuracy. Conclusions: Gafchromic EBT3 film demonstrates reliable dosimetric measurement in high-dose rate brachytherapy. The film's accuracy in evaluating the ${{\rm{\;}}^{192}}$ Ir Flexisource ensures precise treatment delivery and enhances patient safety, supporting its continued use in clinical practice. [ABSTRACT FROM AUTHOR]
- Published
- 2025
- Full Text
- View/download PDF