Background: We evaluated whether the best‐fit intraocular lens (IOL) power formula for the first operated eye (BF1) was also the most accurate formula for the second eye. Methods: This was a retrospective study of 152 patients who underwent uncomplicated delayed bilateral cataract surgery with a minimum delay of 3 weeks using only one 1‐piece IOL (HOYA, Vivinex) at the Medical University of Vienna, Austria. Seven different formulae (Barrett Universal II, Castrop, Haigis, Hoffer Q, Holladay 1, Kane, and SRK/T) were investigated to test the formula selection approach with regard to the calculated mean and median absolute prediction errors (MAE/MedAE). Results: The mean intraindividual difference in axial length was 0.2 mm (±0.3 mm). BF1 coincided with the best‐fit formula for the second eye (BF2) in 56% of patients (p < 0.05). Using BF1 for the second eye led to a lower MedAE (0.22 dioptre, D) than using a formula at random (0.33 D) and was less accurate than using the best‐fit formula for each eye separately (0.1 D). The MedAEs of all formulae were generally low, ranging from 0.28 to 0.35 D. Conclusion: Using BF1 for the second eye led to a lower MedAE than the random selection of a formula. Therefore, BF1 can be used for the second eye if the surgeon is unsure of the choice of formula. [ABSTRACT FROM AUTHOR]