1. Short-term reproducibility of computed tomography-based lung density measurements in alpha-1 antitrypsin deficiency and smokers with emphysema.
- Author
-
Shaker SB, Dirksen A, Laursen LC, Maltbaek N, Christensen L, Sander U, Seersholm N, Skovgaard LT, Nielsen L, and Kok-Jensen A
- Subjects
- Absorptiometry, Photon statistics & numerical data, Aged, Algorithms, Female, Follow-Up Studies, Humans, Image Processing, Computer-Assisted methods, Image Processing, Computer-Assisted statistics & numerical data, Lung diagnostic imaging, Lung physiopathology, Male, Middle Aged, Pulmonary Diffusing Capacity physiology, Pulmonary Emphysema physiopathology, Radiation Dosage, Regression Analysis, Reproducibility of Results, Residual Volume physiology, Software, Tomography, X-Ray Computed statistics & numerical data, Total Lung Capacity physiology, Lung Volume Measurements methods, Pulmonary Emphysema diagnostic imaging, Smoking adverse effects, Tomography, X-Ray Computed methods, alpha 1-Antitrypsin Deficiency complications
- Abstract
Purpose: To study the short-term reproducibility of lung density measurements by multi-slice computed tomography (CT) using three different radiation doses and three reconstruction algorithms., Material and Methods: Twenty-five patients with smoker's emphysema and 25 patients with alpha1-antitrypsin deficiency underwent 3 scans at 2-week intervals. Low-dose protocol was applied, and images were reconstructed with bone, detail, and soft algorithms. Total lung volume (TLV), 15th percentile density (PD-15), and relative area at -910 Hounsfield units (RA-910) were obtained from the images using Pulmo-CMS software. Reproducibility of PD-15 and RA-910 and the influence of radiation dose, reconstruction algorithm, and type of emphysema were then analysed., Results: The overall coefficient of variation of volume adjusted PD-15 for all combinations of radiation dose and reconstruction algorithm was 3.7%. The overall standard deviation of volume-adjusted RA-910 was 1.7% (corresponding to a coefficient of variation of 6.8%). Radiation dose, reconstruction algorithm, and type of emphysema had no significant influence on the reproducibility of PD-15 and RA-910. However, bone algorithm and very low radiation dose result in overestimation of the extent of emphysema., Conclusion: Lung density measurement by CT is a sensitive marker for quantitating both subtypes of emphysema. A CT-protocol with radiation dose down to 16 mAs and soft or detail reconstruction algorithm is recommended.
- Published
- 2004
- Full Text
- View/download PDF