1. T 1ρ for Radiotherapy Treatment Response Monitoring in Rectal Cancer Patients: A Pilot Study.
- Author
-
Kooreman, Ernst S., Tanaka, Max, ter Beek, Leon C., Peters, Femke P., Marijnen, Corrie A. M., van der Heide, Uulke A., and van Houdt, Petra J.
- Subjects
- *
RECTAL cancer , *CANCER patients , *PILOT projects , *LINEAR accelerators , *RADIOTHERAPY - Abstract
Quantitative MRI has the potential to produce imaging biomarkers for the prediction of early response to radiotherapy treatment. In this pilot study, a potential imaging biomarker, the T1ρ relaxation time, is assessed for this purpose. A T1ρ sequence was implemented on a 1.5 T MR-linac system, a system that combines an MRI with a linear accelerator for radiation treatment. An agar phantom with concentrations of 1–4% w/w was constructed for technical validation of the sequence. Phantom images were assessed in terms of short-term repeatability and signal-to-noise ratio. Twelve rectal cancer patients, who were treated with 5 × 5 Gy, were imaged on each treatment fraction. Individual changes in the T1ρ values of the gross tumor volume (GTV) showed an increase for most patients, although a paired t-test comparing values in the GTV from the first to the last treatment fraction showed no statistically significant difference. The phantom measurements showed excellent short-term repeatability (0.5–1.5 ms), and phantom T1ρ values corresponded to the literature values. T1ρ imaging was implemented successfully on the MR-linac, with a repeatability comparable to diagnostic systems, although clinical benefit in terms of treatment response monitoring remains to be demonstrated. [ABSTRACT FROM AUTHOR]
- Published
- 2022
- Full Text
- View/download PDF