1. Compliance Assessment of the Epithelial or Absorbed Power Density Below 10 GHz Using SAR Measurement Systems.
- Author
-
Samaras T, Christ A, and Kuster N
- Subjects
- Algorithms, Phantoms, Imaging, Electromagnetic Fields, Radiometry
- Abstract
The introduction of new dosimetric quantities, in particular, epithelial or absorbed power density for frequencies above 6 GHz, in exposure guidelines and safety standards requires the development of new experimental assessment procedures for compliance testing. In this study, we propose to approximate the peak spatial-average absorbed power density (psS
ab ) using the same measured data and algorithms that are used for determining the peak spatial-average specific absorption rate psSAR, which is currently limited to frequencies up to 10 GHz. The uncertainty component for the transformation of psSAR to psSab was evaluated as less than 0.55 dB (13.5%) for any source as close as 0.02 λ from the tissue simulating media. The approach is easy to implement and allows determining compliance with the basic restrictions of the latest safety guidelines. In the next project, we will expand dosimetric probes, phantoms, and procedures for frequencies above 10 GHz. © 2021 Bioelectromagnetics Society., (© 2021 Bioelectromagnetics Society.)- Published
- 2021
- Full Text