1. Correcting systematic error in PO 2 measurement to improve measures of oxygen supply capacity (α).
- Author
-
Timpe AW and Seibel BA
- Subjects
- Calibration, Animals, Partial Pressure, Hypoxia metabolism, Oxygen metabolism, Oxygen Consumption
- Abstract
An organism's oxygen supply capacity, measured as a ratio of a metabolic rate to its critical oxygen partial pressure, describes the efficacy of oxygen uptake and transport. This metric is sensitive to errors in oxygen measurement, especially near anoxia where the magnitude of instrument error as a proportion of total signal is magnified. Here, we present a conceptual and mathematical method that uses this sensitivity to identify, quantify, and therefore correct oxygen measurements collected using inaccurately calibrated sensors. When appropriate, adding a small correction value to each oxygen measurement counteracts the effects of this error and provides results that are comparable to data from accurately calibrated oxygen probes. We demonstrate, using simulated, laboratory, and literature datasets, how this method can be used post hoc to diagnose error in, correct the magnitude of, and reduce the variability in repeat measures of traits relevant to oxygen tolerance., Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper., (Copyright © 2024. Published by Elsevier Inc.)
- Published
- 2024
- Full Text
- View/download PDF