1. A real-world study evaluating ultrasound-guided percutaneous non-targeted liver biopsy needle failures and pathology sample-quality assessment in both end-cut and side-notch needles.
- Author
-
Zhong J, Allard R, Hewitson D, Weston M, Hulson O, Burbidge S, Lambie H, Kaye T, Wyatt JI, and Albazaz R
- Subjects
- Adolescent, Adult, Aged, Aged, 80 and over, Biopsy, Needle instrumentation, Biopsy, Needle methods, Equipment Design, Female, Humans, Image-Guided Biopsy instrumentation, Image-Guided Biopsy methods, Liver diagnostic imaging, Male, Middle Aged, Reproducibility of Results, Young Adult, Equipment Failure statistics & numerical data, Liver pathology, Ultrasonography, Interventional methods
- Abstract
Objectives: To determine biopsy device failures, causative factors, complications and sample quality of the 16G end-cut Biopince
™ and side-notch Bard™ needles., Methods: All ultrasound-guided non-targeted liver biopsies between 01/01/2016 and 31/12/2018 were included. Operator, device, number of failures, complications and repeat biopsies were recorded. Histopathology samples were reviewed for all cases of needle failure and a group with no failures, and graded "yes/no" for the presence of steatosis, inflammation and fibrosis. The pathology slides from these cases were reviewed to assess biopsy sample quality (length and portal tract number). The failure and no-failure groups were compared in terms of device type/histology, and sample quality was compared between the needle types., Results: 1004 patients were included. 93.8% ( n = 942) required one needle pass to obtain a sample and 6.2% ( n = 62) required >1 pass due to needle failure. Total of 76 needle failures, more with end-cut than side-notch needles (8.7% vs 2.9%) ( p < 0.001). No needle failures resulted in complication. The presence of liver fibrosis was associated with fewer needle failures ( p = 0.036). The major complication rate was 0.4% (4/1044). A biopsy with >10 portal tracts was obtained in 90.2% of specimens > 20 mm long, compared with 66% of 16-20 mm biopsies and 21% of <16 mm biopsies. The target of >10 portal tracts was achieved in 10/26 (38.5%) of side-notch biopsies and 64/90 (71.1%) of end-cut biopsies ( p = 0.004)., Conclusion: Ultrasound-guided liver biopsy is safe and sample quality is consistently good when a core >20 mm long is obtained. The end-cut biopsy device generated reliably good quality biopsy samples; however, the needle failure rate was significantly higher than the side-cut needle., Advances in Knowledge: Ultrasound-guided liver biopsy specimen quality is consistently good when a core >20 mm long is obtained which can be achieved with a single pass using the 16G BiopinceTM end-cut needle, although the needle failure rate is significantly higher than the 16G Max-Core™ Bard™ side-notch needle.- Published
- 2021
- Full Text
- View/download PDF