1. Efficacy and Safety of Gastrointestinal Tumour Site Marking with da Vinci-Compatible Near-Infrared Fluorescent Clips: A Case Series.
- Author
-
Takahashi J, Yoshida M, Ohdaira H, Nakaseko Y, Nakashima K, Kamada T, Suzuki N, Sato T, and Suzuki Y
- Subjects
- Humans, Prospective Studies, Rectum, Surgical Instruments, Coloring Agents, Gastrointestinal Neoplasms diagnostic imaging, Gastrointestinal Neoplasms surgery, Laparoscopy methods, Robotic Surgical Procedures methods
- Abstract
Background: The conventional near-infrared fluorescent clip (NIRFC) ZEOCLIP FS® has been used successfully in marking tumour sites during laparoscopic surgeries. However, this clip is difficult to observe with the Firefly imaging system equipped with the da Vinci® surgical system. We have been involved in the modification of ZEOCLIP FS® and development of da Vinci-compatible NIRFC. This is the first prospective single-centre case series study verifying the usefulness and safety of the da Vinci-compatible NIRFC., Methods: Twenty-eight consecutive patients undergoing da Vinci®-assisted surgery for gastrointestinal cancer (16 gastric, 4 oesophageal, and 8 rectal cases) between May 2021 and May 2022 were enrolled., Results: Tumour location was identified by the da Vinci-compatible NIRFCs in 21 of 28 (75%) patients, which involved 12 gastric (75%), 4 oesophageal (100%), and 5 rectal (62%) cancer cases. No adverse events were observed., Conclusion: Tumour site marking with da Vinci-compatible NIRFC was feasible in 28 patients enrolled in this study. Further studies are warranted to substantiate the safety and improve the recognition rate., (© 2023. The Author(s).)
- Published
- 2023