1. A general method for quantitative measurement of molecular mass distribution by mass spectrometry.
- Author
-
Park ES, Wallace WE, Guttman CM, Flynn KM, Richardson MC, and Holmes GA
- Subjects
- Computer Simulation, Algorithms, Fullerenes analysis, Fullerenes chemistry, Models, Chemical, Molecular Weight, Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization methods
- Abstract
A method is presented to test whether the conversion of the mass spectrum of a polydisperse analyte to its molecular mass distribution is quantitative. Mixtures of samples with different average molecular masses, coupled with a Taylor's expansion mathematical formalism, were used to ascertain the reliability of molecular mass distributions derived from mass spectra. Additionally, the method describes how the molecular mass distributions may be corrected if the degree of mass bias is within certain defined limits. This method was demonstrated on polydisperse samples of C(60) fullerenes functionalized with ethylpyrrolidine groups measured by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry; however, it is applicable to any polydisperse analyte and mass spectrometric method as long as spectrum resolution allows individual oligomers to be identified. Mass spectra of the derivatized fullerenes taken in positive ion mode were shown to give an accurate measurement of the molecular mass distribution while those taken in negative ion mode were not. Differences in the mechanisms for ion formation are used to explain the discrepancy. Official contribution of the National Institute of Standards and Technology; not subject to copyright in the United States of America.
- Published
- 2009