1. Rise-Time Calibration of 50-GHz Sampling Oscilloscopes: Intercomparison Between PTB and NPL.
- Author
-
Bieler, Mark, Seitz, Steffen, Spitzer, Meinhard, Hein, Günter, Pierz, Klaus, Siegner, Uwe, Basu, M. Anjan, Smith, Andrew J. A., and Harper, Matthew R.
- Subjects
CALIBRATION ,TESTING laboratories ,ELECTROOPTICS ,INTEGRATED optics ,DIGITAL oscilloscopes ,OSCILLOSCOPES ,ELECTRONIC instruments - Abstract
We present the results of the first intercomparison of the rise-time calibration of a 50-GHz sampling oscilloscope between Physikalisch-Technische Bundesanstalt, Germany, and National Physical Laboratory, U.K. (EUROMET project 641). Rise-times between 6.0 and 6.6 ps were obtained using optoelectronic techniques. Detailed uncertainty budgets are given, which yield typical rise-time uncertainties of 1.2 ps. The absolute values of the degree of equivalence are ≤ 0.3 ps with an expanded uncertainty of 1.0 ps and, thus, validate the experimental techniques. [ABSTRACT FROM AUTHOR]
- Published
- 2007
- Full Text