1. End tidal carbon dioxide is as reliable as transcutaneous monitoring in ventilated postsurgical neonates.
- Author
-
Tingay DG, Mun KS, and Perkins EJ
- Subjects
- Blood Gas Monitoring, Transcutaneous methods, Capnography methods, Carbon Dioxide blood, Feasibility Studies, Humans, Infant Care methods, Infant, Newborn, Monitoring, Physiologic methods, Partial Pressure, Reproducibility of Results, Respiration, Artificial, Tidal Volume, Carbon Dioxide analysis, Postoperative Care methods
- Abstract
Objectives: To compare the agreement, precision and repeatability of end tidal carbon dioxide (EtCO2 ) and transcutaneous carbon dioxide (TcCO2 ) with partial pressure of arterial CO(2) ( PaCO2) in postoperative neonates., Patients: Fifty mechanically ventilated neonates without lung disease, and with no contraindications for either TcCO2 or EtCO2 monitoring., Interventions: Paired TcCO2 and EtCO2 values were recorded with three consecutive measurements within the first 48 h of surgery., Main Outcome Measures: EtCO2, TcCO2 and PaCO2 triplets were compared using Bland-Altman plots., Results: One hundred thirty-two triplet measures of CO(2) were recorded with mean PaCO2 43.5 (7.3) mm Hg, EtCO2 38.8 (6.4) mm Hg and 43.8 (8.8) mm Hg (p<0.0001 for EtCO2 against PaCO2; paired t test). The PaCO2 - EtCO2 bias±2SD was 4.1±9.0 mm Hg and -0.8±13.0 mm Hg for PaCO2 - TcCO2. 56.1% of EtCO2, and 60.6% of TcCO2 values were within ±5 mm Hg of paired PaCO2., Conclusions: In postoperative neonates, EtCO2 and TcCO2 demonstrated a clinically acceptable agreement with PaCO2.
- Published
- 2013