1. A novel true volumetric method for the determination of residual leucocytes in blood components
- Author
-
K. Tapernon, Uwe Cassens, Burkhard Greve, Walter Sibrowski, B. Nave, Wolfgang Göhde, and Erhard Severin
- Subjects
Detection limit ,medicine.medical_specialty ,Chromatography ,business.industry ,Fluorescence spectrometry ,Automatic processing ,Hematology ,General Medicine ,Residual ,Surgery ,Red blood cell ,medicine.anatomical_structure ,hemic and lymphatic diseases ,Blood Component Transfusion ,Cost control ,Medicine ,Platelet concentrate ,business ,circulatory and respiratory physiology - Abstract
BACKGROUND AND OBJECTIVES Accurate determination of residual leucocytes [white blood cells (WBC)] in blood components is of high clinical importance. To date, several labour-intensive, time-consuming or expensive techniques have been used for this purpose. MATERIALS AND METHODS A method for the determination of residual WBC is described using a novel low-cost flow-cytometric cell counter and analyser (CCA). The DNA in WBC was stained using 4'-6-diamidino-2-phenylindole (DAPI) and WBC were automatically analysed by true volumetric counting of 200-microl samples (prepared from a 20-microl undiluted sample). RESULTS Dilution experiments over a range of 0.5-50 WBC/microl showed a linearity of r = 0.998. The detection limit of this method was 0.83 WBC/microl of red blood cell concentrate (RCC) and 0.67 WBC/microl of platelet concentrate (PC), with an accuracy of 95.5%. CONCLUSION Residual WBC (< 1 WBC/microl) can be accurately counted using the CCA within 2 min and at a total cost of less than euro 1 per sample.