1. Clinical experience with a multidetector SPET system (Toshiba GCA-9300A).
- Author
-
Mahmood S, Buscombe JR, Kouris K, Clarke GA, Townsend CE, Jarritt PH, Costa DC, and Ell PJ
- Subjects
- Bone and Bones diagnostic imaging, Humans, Kidney diagnostic imaging, Male, Middle Aged, Organotechnetium Compounds, Oximes, Reference Values, Sensitivity and Specificity, Technetium Tc 99m Exametazime, Tomography, Emission-Computed, Single-Photon methods, Brain diagnostic imaging, Brain Diseases diagnostic imaging, Chest Pain diagnostic imaging, Myocardial Ischemia diagnostic imaging, Tomography, Emission-Computed, Single-Photon instrumentation
- Abstract
The clinical experience with the Toshiba GCA-9300A single photon emission tomography (SPET) system is discussed along with typical acquisition protocols for various SPET studies. The system was used to perform SPET studies in normals and in a variety of brain and body disorders. Its three Anger-type gamma cameras forming a triangular aperture offer a substantial increase in sensitivity compared to a single rotating gamma camera. This has allowed the routine use of lead fanbeam super high-resolution collimators (SHR FB) for 99Tcm-hexamethylpropyleneamine oxime (HMPAO) brain SPET studies and high-resolution parallel-hole collimators (HR PH) for cardiac and other body studies. The resulting improvement in spatial resolution coupled with the ease of patient positioning and the greater patient throughput compared to a conventional tomographic gamma camera, will enhance the role of brain and body SPET for both routine and research purposes.
- Published
- 1994
- Full Text