1. Very long-term outcomes of the Carpentier-Edwards Perimount aortic valve in patients aged 50-65 years.
- Author
-
Bourguignon T, Lhommet P, El Khoury R, Candolfi P, Loardi C, Mirza A, Boulanger-Lothion J, Bouquiaux-Stablo-Duncan AL, Marchand M, and Aupart M
- Subjects
- Aged, Female, Heart Valve Diseases surgery, Humans, Kaplan-Meier Estimate, Male, Middle Aged, Retrospective Studies, Aortic Valve surgery, Bioprosthesis, Heart Valve Prosthesis, Heart Valve Prosthesis Implantation adverse effects, Heart Valve Prosthesis Implantation instrumentation, Heart Valve Prosthesis Implantation mortality
- Abstract
Objectives: Aortic valve replacement (AVR) using a bioprosthesis remains controversial for patients aged 50-65 years. This cohort study reports the very long-term outcomes of AVR using Carpentier-Edwards Perimount pericardial bioprosthesis in this age group., Methods: From 1984 to 2008, 522 Carpentier-Edwards Perimount pericardial aortic bioprostheses were implanted in 516 patients aged 50-65 years (mean age, 60 ± 4 years; 19% female). Multiple valve replacements were excluded fro m our cohort. Baseline demographic, perioperative and follow-up data were recorded prospectively. Mean follow-up was 9 ± 6 years, for a total of 4428 valve-years. Follow-up was complete for 97% of patients included., Results: Operative mortality rate was 2%. One hundred and forty-six late deaths occurred for a linearized rate of 3%/valve-year. Actuarial survival rates averaged 73 ± 2, 59 ± 3 and 35 ± 5% after 10, 15 and 20 years of follow-up, respectively. Mortality rate associated with reoperation was 2%. Actuarial freedom from reoperation rates due to structural valve deterioration (SVD) at 10, 15 and 20 years was respectively of 91 ± 2, 76 ± 3 and 50 ± 6%. Competing risk analysis demonstrated an actual risk of explantation secondary to SVD at 20 years of 30 ± 3%. Expected valve durability was 19 years for this age group. Age was not a significant risk factor for SVD in this middle-aged population., Conclusions: In patients aged 50-65 years undergoing AVR with the Carpentier-Edwards Perimount bioprosthesis, the expected valve durability was 19 years. Age was not a significant risk factor for SVD within this age group. Patient selection and attention to timing of reintervention may be determinants of long-term outcomes., (© The Author 2015. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.)
- Published
- 2016
- Full Text
- View/download PDF