1. The Fabrication and Sensing Characteristics of Arrayed Flexible IGZO/Al Urea Biosensor Modified by Graphene Oxide
- Author
-
Hong-Yu Huang, You-Xiang Wu, Jung-Chuan Chou, Chih-Hsien Lai, Cian-Yi Wu, Siao-Jie Yan, and Yi-Hung Liao
- Subjects
Indium gallium zinc oxide ,Materials science ,Urease ,biology ,Graphene ,010401 analytical chemistry ,Inorganic chemistry ,Oxide ,Substrate (chemistry) ,02 engineering and technology ,Buffer solution ,021001 nanoscience & nanotechnology ,01 natural sciences ,0104 chemical sciences ,Computer Science Applications ,law.invention ,chemistry.chemical_compound ,chemistry ,Chemical engineering ,law ,Polyethylene terephthalate ,biology.protein ,Electrical and Electronic Engineering ,0210 nano-technology ,Biosensor - Abstract
The urease membrane was immobilized on the surface of indium gallium zinc oxide (IGZO) based on polyethylene terephthalate substrate, which was used to develop an arrayed flexible urea biosensor in this study. In order to improve the properties of arrayed flexible IGZO/Al urea biosensor, graphene oxide (GO) was employed to modify the properties of arrayed flexible IGZO urea biosensor. Compared with the average sensitivity of 1.53 mV/(mg/dl) and linearity of 0.931 for arrayed flexible IGZO/Al urea biosensor, the average sensitivity and linearity of arrayed flexible IGZO urea biosensor modified by GO achieved 3.06 mV/(mg/dl) and 0.970, respectively. As the result, 5 mM pH 7.0 PBS buffer solution was the optimal measurement environment for Urease/IGZO/GO/Al urea biosensor, and which the average sensitivity was 3.65 [mV/(mg/dl)]. Anti-interference testing showed the arrayed flexible IGZO/Al urea biosensor modified by GO was not easily interfered from the other materials.
- Published
- 2017