1. Adiabatic inversion pulses for myocardial T1 mapping.
- Author
-
Kellman P, Herzka DA, and Hansen MS
- Subjects
- Humans, Phantoms, Imaging, Reproducibility of Results, Sensitivity and Specificity, Algorithms, Heart anatomy & histology, Image Enhancement methods, Image Interpretation, Computer-Assisted methods, Magnetic Resonance Imaging methods
- Abstract
Purpose: To evaluate the error in T1 estimates using inversion-recovery-based T1 mapping due to imperfect inversion and to perform a systematic study of adiabatic inversion pulse designs in order to maximize inversion efficiency for values of transverse relaxation (T2) in the myocardium subject to a peak power constraint., Methods: The inversion factor for hyperbolic secant and tangent/hyperbolic tangent adiabatic full passage waveforms was calculated using Bloch equations. A brute-force search was conducted for design parameters: pulse duration, frequency range, shape parameters, and peak amplitude. A design was selected that maximized the inversion factor over a specified range of amplitude and off-resonance and validated using phantom measurements. Empirical correction for imperfect inversion was performed., Results: The tangent/hyperbolic tangent adiabatic pulse was found to outperform hyperbolic secant designs and achieve an inversion factor of 0.96 within ±150 Hz over 25% amplitude range with 14.7 µT peak amplitude. T1 mapping errors of the selected design due to imperfect inversion was ∼4% and could be corrected to <1%., Conclusions: Nonideal inversion leads to significant errors in inversion-recovery-based T1 mapping. The inversion efficiency of adiabatic pulses is sensitive to transverse relaxation. The tangent/hyperbolic tangent design achieved the best performance subject to the peak amplitude constraint., (Copyright © 2013 Wiley Periodicals, Inc.)
- Published
- 2014
- Full Text
- View/download PDF