1. Silagerm 8040-Based Planar Microfluidic Chips for Sampling and Sample Preparation in the Analysis of Gas Media.
- Author
-
Margaryan, A. E., Platonov, I. A., Kolesnichenko, I. N., Novikova, E. A., and Karsunkina, A. S.
- Subjects
- *
AIR analysis , *AIR sampling , *GAS analysis , *DIABETES , *DESORPTION - Abstract
A technology is developed for manufacturing planar microfluidic chips (MFCs) based on Silgard 8040, encompassing template fabrication, mold production, sealing, and adsorbent filling. These MFCs, filled with a Porapak-Q adsorbent, facilitate sample preparation and calibration under uniform conditions, integrating both sample collection and preconcentration stages. Experimental investigations have identified optimal conditions for acetone preconcentration using Silgard 8040-based MFCs, achieving a maximum concentration factor of 43 at adsorption temperature (Tads) of 0°C, desorption temperature (Tdes) of 70°C, adsorption volume (Vads) of 45 mL, and desorption time (tdes) of 1 s. The feasibility of using MFCs for sample collection and preparation, taking into account sample storage time, has been evaluated. The procedure maintains accuracy within 6–10% over 8 h without a necessity for additional drying of the collected samples from exhaled air. [ABSTRACT FROM AUTHOR]
- Published
- 2024
- Full Text