1. Hip joint centre localisation with an unscented Kalman filter.
- Author
-
Momi, Elena De, Beretta, Elisa, and Ferrigno, Giancarlo
- Subjects
- *
HIP joint , *KALMAN filtering , *STOCHASTIC processes , *ANATOMY , *PRESERVATION of organs, tissues, etc. - Abstract
The accurate estimation of the hip joint centre (HJC) in gait analysis and in computer assisted orthopaedic procedures is a basic requirement. Functional methods, based on rigid body localisation, assessing the kinematics of the femur during circumduction movements (pivoting) have been used for estimating the HJC. Localising the femoral segment only, as it is usually done in total knee replacement procedure, can give rise to estimation errors, since the pelvis, during the passive pivoting manoeuvre, might undergo spatial displacements. This paper presents the design and test of an unscented Kalman filter that allows the estimation of the HJC by observing the pose of the femur and the 3D coordinates of a single marker attached to the pelvis. This new approach was validated using a hip joint mechanical simulator, mimicking both hard and soft tissues. The algorithm performances were compared with the literature standards and proved to have better performances in case of pelvis translation greater than 8 mm, thus satisfying the clinical requirements of the application. [ABSTRACT FROM AUTHOR]
- Published
- 2013
- Full Text
- View/download PDF