1. High defibrillation threshold with right-sided ICD implantation was resolved by a dual coil lead via persistent left superior vena cava
- Author
-
Naoyuki Kamiyama, Tetsuji Miura, Daigo Nagahara, Atsushi Mochizuki, Taro Tsuzuki, and Takefumi Fujito
- Subjects
medicine.medical_specialty ,Defibrillation ,business.industry ,medicine.medical_treatment ,Case Report ,Ventricular tachycardia ,medicine.disease ,Implantable cardioverter-defibrillator ,Defibrillation threshold ,Internal medicine ,Idiopathic dilated cardiomyopathy ,Ventricular fibrillation ,medicine ,Cardiology ,Persistent left superior vena cava ,Cardiology and Cardiovascular Medicine ,business ,Coronary sinus - Abstract
Persistent left superior vena cava (PLSVC) can be problematic when device implantation is scheduled from the left side because of the technical difficulty in delivering leads. Right-sided implantation is an alternative method, but there is a risk of a high defibrillation threshold (DFT). Transvenous implantation of an implantable cardioverter defibrillator (ICD) was scheduled for a 54-year-old man with idiopathic dilated cardiomyopathy and monomorphic non-sustained ventricular tachycardia, but computed tomography revealed the presence of a PLSVC. Right-sided ICD implantation was performed first; however, an ICD shock at 35 J failed to terminate the induced ventricular fibrillation (VF). Re-implantation via the PLSVC by a left subclavian approach with a dual coil lead was performed next. The dual coil right ventricular lead was successfully implanted via the PLSVC, and the induced VF was terminated by a single shock at 25 J. In the present case, the proximal coil was located in the coronary sinus (CS) and it enabled an antero-posterior defibrillation vector across the left ventricle. In addition to the re-location of the ICD generator from the right side to the left side, the new positioning of the proximal coil inside the CS is likely to have contributed to the great improvement of the DFT.
- Published
- 2021