1. Monte Carlo model of the scanning beam digital x-ray (SBDX) source.
- Author
-
Bazalova, M., Weil, M. D., Wilfley, B., and Graves, E. E.
- Subjects
MONTE Carlo method ,FLUOROSCOPY ,COLLIMATORS ,IMAGING systems ,THIN films ,ELECTRIC potential ,DRUG dosage - Abstract
The scanning-beam digital x-ray (SBDX) system has been developed for fluoroscopic imaging using an inverse x-ray imaging geometry. The SBDX system consists of a large-area x-ray source with a multihole collimator and a small detector. The goal of this study was to build a Monte Carlo (MC) model of the SBDX source as a useful tool for optimization of the SBDX imaging system in terms of its hardware components and imaging parameters. The MC model of the source was built in the EGSnrc/BEAMnrc code and validated using the DOSXYZnrc code and Gafchromic film measurements for 80, 100, and 120 kV x-ray source voltages. The MC simulated depth dose curves agreed with measurements to within 5%, and beam profiles at three selected depths generally agreed within 5%. Exposure rates and half-value layers for three voltages were also calculated from the MC simulations. Patient skin-dose per unit detector-dose was quantified as a function of patient size for all three x-ray source voltages. The skin-dose to detector-dose ratio ranged from 5–10 for a 20 cm thick patient to 1 × 10
3 –1 × 105 for a 50 cm patient for the 120 and 80 kV beams, respectively. Simulations of imaging dose for a prostate patient using common imaging parameters revealed that skin-dose per frame was as low as 0.2 mGy. [ABSTRACT FROM AUTHOR]- Published
- 2012