D'Mello, Yannick, Skoric, James, Xu, Shicheng, Akhras, Megan, Roche, Philip J. R., Lortie, Michel A., Gagnon, Stephane, and Plant, David V.
We present a novel seismocardiography (SCG)-based approach for real-time cardio-respiratory activity measurement called the Autocorrelated Differential Algorithm (ADA). Measurements were performed on ten male subjects in the supine position for three 7-minute-long sets each, corresponding to 14,619 heartbeats. The ADA utilized temporal variations, windowing, and autocorrelation to produce physiological measurements corresponding to heart rate (HR), and left ventricular ejection time, and estimations of respiration rate, volume, and phase. The versatility of the ADA was investigated in two contexts: physical exertion and heart rate variability. The accuracy of HR measurements at a sampling frequency of 200 Hz resulted in a correlation coefficient ($r^{2}$) of 0.9808 when compared with a manual annotation of all datasets. Its reproducibility was tested on externally obtained SCG and electrocardiography datasets, which produced an $r^{2}$ of 0.8224. The accuracy and computational time were also characterized by different sampling frequencies to quantify performance. The recommended sampling frequency is 200 Hz corresponding to a computation time of 0.05 s per instantaneous measurement using a standard desktop computer. The ADA delivered real-time SCG measurements with a refresh rate that was dependent on the computational time per measurement, which could be decreased by lowering the sampling frequency. The presented algorithm offers a novel tool toward real-time physiological monitoring in clinical and everyday scenarios. [ABSTRACT FROM AUTHOR]