1. A case of total hip arthroplasty for nonunion after femoral trochanteric fracture surgery with complete breakage of only the compression screw of the InterTAN nail
- Author
-
Fumihiro Yoshimura, Koichi Kinoshita, Hajime Seo, Taiki Matsunaga, Kenichiro Doi, and Takuaki Yamamoto
- Subjects
Trigen InterTAN nail ,Mechanical failure ,Nail ,Lag screw ,Total hip arthroplasty ,Surgery ,RD1-811 - Abstract
Intramedullary nails are an effective treatment for common femoral trochanteric fractures. However, one of their complications is implant breakage due to poor reduction and nonunion after surgery. We herein report a case of a 54-year-old man who underwent total hip arthroplasty for nonunion after internal fixation of a femoral trochanteric fracture. The femoral trochanteric fracture was treated by internal fixation using the Trigen InterTAN nail. The patient developed symptoms of hip pain 6 months after internal fixation. Nine months after internal fixation, hip radiographs and computed tomography scans showed breakage of only the compression screw. During total hip arthroplasty, we were unable to remove the lag screw and compression screw before the femoral head dislocation because no gap was present between the two screws. Thus, we removed these screws with the femoral head after dislocation of the femoral head. The removed nail was partially damaged at the lag screw hole. This change was retrospectively observed on the preoperative computed tomography scan. Three months after total hip arthroplasty, the patient was able to walk unaided and the hip pain had resolved. If only the compression screw is completely broken after internal fixation with the Trigen InterTAN nail, both the lag screw and compression screw will be difficult to remove with preservation of the femoral head. We effectively managed such a case by not only revision internal fixation but also total hip arthroplasty.
- Published
- 2023
- Full Text
- View/download PDF