1. Comparison of the force applied on oral structures during intubation attempts by novice physicians between the Macintosh direct laryngoscope, Airway Scope and C-MAC PM: a high-fidelity simulator-based study.
- Author
-
Taizo Nakanishi, Takashi Shiga, Yosuke Homma, Yasuaki Koyama, and Tadahiro Goto
- Abstract
Objective: We examined whether the use of Airway Scope (AWS) and C-MAC PM (C-MAC) decreased the force applied on oral structures during intubation attempts as compared with the force applied with the use of Macintosh direct laryngoscope (DL). Design: Prospective cross-over study. Participants: A total of 35 novice physicians participated. Interventions: We used 6 simulation scenarios based on the difficulty of intubation and intubation devices. Outcome measures: Our primary outcome measures were the maximum force applied on the maxillary incisors and tongue during intubation attempts, measured by a high-fidelity simulator. Results: The maximum force applied on maxillary incisors was higher with the use of the C-MAC than with the DL and AWS in the normal airway scenario (DL, 26 Newton (N); AWS, 18 N; C-MAC, 52 N; p<0.01) and the difficult airway scenario (DL, 42 N; AWS, 24 N; C-MAC, 68 N; p<0.01). In contrast, the maximum force applied on the tongue was higher with the use of the DL than with the AWS and C-MAC in both airway scenarios (DL, 16 N; AWS, 1 N; C-MAC, 7 N; p<0.01 in the normal airway scenario; DL, 12 N; AWS, 4 N; CMAC, 7 N; p<0.01 in the difficult airway scenario). Conclusions: The use of C-MAC, compared with the DL and AWS, was associated with the higher maximum force applied on maxillary incisors during intubation attempts. In contrast, the use of video laryngoscopes was associated with the lower force applied on the tongue in both airway scenarios, compared with the DL. Our study was a simulationbased study, and further research on living patients would be warranted. [ABSTRACT FROM AUTHOR]
- Published
- 2016
- Full Text
- View/download PDF