1. Evaluation of a Fourth-Generation Subcutaneous Real-Time Continuous Glucose Monitor (CGM) in Individuals With Diabetes on Peritoneal Dialysis
- Author
-
Jack K.C. Ng, James Ling, Andrea O.Y. Luk, Eric S.H. Lau, Ronald C.W. Ma, Philip K.T. Li, Cheuk Chun Szeto, Juliana C.N. Chan, and Elaine Chow
- Subjects
Advanced and Specialized Nursing ,Endocrinology, Diabetes and Metabolism ,Internal Medicine - Abstract
OBJECTIVE To evaluate the performance of a real-time continuous glucose monitor (CGM) in individuals with diabetes on peritoneal dialysis (PD). RESEARCH DESIGN AND METHODS Thirty participants with type 2 diabetes on continuous ambulatory PD wore a Guardian Sensor 3 on the upper arm paired with Guardian Connect for 14 days. We compared CGM readings against Yellow Springs Instrument (YSI) venous glucose during an 8-h in-clinic session with glucose challenge. RESULTS The mean absolute relative difference (MARD) was 10.4% (95% CI 9.6, 11.7) from 941 CGM-YSI matched pairs; 81.3% of readings were within %15/15 of YSI values in the full glycemic range. Consensus error grid analysis showed 99.9% of sensor values in zones A and B. There were no correlations between pH, uremia, hydration status, and MARD. CONCLUSIONS We showed satisfactory performance of a real-time CGM sensor in PD patients with diabetes, supporting future use to facilitate treatment decisions.
- Published
- 2023
- Full Text
- View/download PDF