1. Increasing the sensitivity of Listeria monocytogenes assays: evaluation using ELISA and amperometric detection
- Author
-
L. Crowley, Elizabeth, K. O’Sullivan, Ciara, and G. Guilbault, George
- Abstract
An immunosensor for the detection of Listeria monocytogenes was developed. ELISA and amperometric studies were run in parallel to develop a more sensitive and rapid assay for the bacterium. Conditions for the immunosensor were primarily characterised using ELISA. A direct sandwich assay was employed and the affinities of two polyclonal (goat and rabbit) and one monoclonal (mouse) anti-L. monocytogenes antibodies were compared using this format. Owing to low sensitivity being obtained with all antibodies, biotin–avidin amplification and an indirect sandwich assay were employed. The system was then transferred to screen-printed electrodes (SPEs), the primary antibody being immobilised by cross-linking with 1-ethyl-3-(3-dimethylaminopropyl)-carbodiimide, and the mode of detection being amperometric. Various parameters (limit of detection, working range, incubation time, cross-reactivity) of the systems were characterised. The effect of direct incubation in milk is also discussed. The final immunosensor had a working range of 1 × 106–1 × 103 cells ml–1 and a detection limit of 9 × 102 cells ml–1. The assay took about 3.5 h to complete.
- Published
- 1999
- Full Text
- View/download PDF