1. A quantitative method for determining a representative detection limit of the forensic luminol test for latent bloodstains.
- Author
-
Cassidy BM, Lu Z, Martin JP, Tazik SK, Kellogg KW, DeJong SA, Belliveau EO, Kilgore KE, Ervin SM, Meece-Rayle M, Abraham AM, Myrick ML, and Morgan SL
- Subjects
- Forensic Medicine methods, Humans, Indicators and Reagents, Limit of Detection, Software, Video Recording, Blood Stains, Luminescent Measurements, Luminol
- Abstract
The luminol test has been used for over 60 years by forensic investigators for presumptive identification of blood and visualization of blood splatter patterns. Multiple studies have estimated the limit of detection (LD) for bloodstains when luminol is employed, with results ranging from 100× to 5,000,000× dilute. However, these studies typically have not identified and controlled important experimental variables which may affect the luminol LD for bloodstains. Without control of experimental parameters in the laboratory, variables which affect the potential of presumptive bloodstain test methods remain largely unknown, and comparisons required to establish new, more powerful detection methods are simply impossible. We have developed a quantitative method to determine the relationship between the amount of blood present and its reaction with luminol by measuring, under controlled conditions, the resulting chemiluminescent intensity with a video camera, combined with processing of the digital intensity data. The method resulted in an estimated LD for bloodstains on cotton fabric at ∼200,000× diluted blood with a specific luminol formulation. Although luminol is the focus of this study, the experimental protocol used could be modified to study effects of variables using other blood detection reagents., (Copyright © 2017 Elsevier B.V. All rights reserved.)
- Published
- 2017
- Full Text
- View/download PDF