1. Validating the Use of Continuous Glucose Monitors With Nondiabetic Recreational Runners.
- Author
-
Mason, Lesley J., Hartwig, Timothy, and Greene, David
- Subjects
PSYCHOLOGY of athletes ,FOOD consumption ,RECREATION ,RUNNING ,RESEARCH methodology evaluation ,DESCRIPTIVE statistics ,CONTINUOUS glucose monitoring ,RESEARCH methodology ,AEROBIC exercises ,OXYGEN consumption ,COMPARATIVE studies - Abstract
Purpose: Continuous glucose monitors (CGMs) are becoming increasingly popular among endurance athletes despite unconfirmed accuracy. We assessed the concurrent validity of the FreeStyle Libre 2 worn on 2 different sites at rest, during steady-state running, and postprandial. Methods: Thirteen nondiabetic, well-trained recreational runners (age = 40 [8] y, maximal aerobic oxygen consumption = 46.1 [6.4] mL·kg
–1 ·min–1 ) wore a CGM on the upper arm and chest while treadmill running for 30, 60, and 90 minutes at intensities corresponding to 50%, 60%, and 70% of maximal aerobic oxygen consumption, respectively. Glucose was measured by manually scanning CGMs and obtaining a finger-prick capillary blood glucose sample. Mean absolute relative difference, time in range, and continuous glucose Clarke error grid analysis were used to compare paired CGM and blood glucose readings. Results: Across all intensities of steady-state running, we found a mean absolute relative difference of 13.8 (10.9) for the arm and 11.4 (9.0) for the chest. The coefficient of variation exceeded 70%. Approximately 47% of arm and 50% of chest paired glucose measurements had an absolute difference ≤10%. Continuous glucose Clarke error grid analysis indicated 99.8% (arm) and 99.6% (chest) CGM data fell in clinically acceptable zones A and B. Time-in-range analysis showed reduced accuracy at lower glucose levels. However, CGMs accurately detected trends in mean glucose readings over time. Conclusions: CGMs are not valid for point glucose monitoring but appear to be valid for monitoring glucose trends during steady-state exercise. Accuracy is similar for arm and chest. Further research is needed to determine whether CGMs can detect important events such as hypoglycemia during exercise. [ABSTRACT FROM AUTHOR]- Published
- 2024
- Full Text
- View/download PDF