1. A potential infection hazard associated with the use of disposable saline vials.
- Author
-
Rutala WA, Stiegel MM, and Sarubbi FA Jr
- Subjects
- Cross Infection prevention & control, Equipment Design, Hand Disinfection, Humans, Intensive Care Units, Skin microbiology, Suction, Trachea, Bacterial Infections etiology, Disposable Equipment statistics & numerical data, Drug Packaging methods, Sodium Chloride administration & dosage
- Abstract
Nosocomial respiratory tract infections have occasionally been associated with contaminated respiratory therapy devices and techniques. In the past two years, our hospital purchased disposable saline squeeze vials for use during suctioning of intubated patients. These vials have a cap which must be flipped or twisted-off by the user before the contents can be instilled into the patient's respiratory tract. We observed use of this item in our intensive care units (ICU) and studied the potential for contamination. ICU nurses use ungloved hand(s) to flip or twist-off squeeze vial cups resulting in skin contact with the vial opening. Cultures of nurses' hands and squeeze vial contents were obtained in the ICU where 24 nurses opened 92 vials in the manner previously described. Twenty-three percent of vial contents were contaminated and cultures revealed S. epidermidis, S. aureus, Streptococcus viridans and enterococcus. On six occasions, organisms recovered from vial contents were believed to be identical to organisms isolated from the hands of the nurse who opened the vial. When nurses used care in removing the saline vial cap, cultures of vial contents were sterile.
- Published
- 1984
- Full Text
- View/download PDF