1. The Tenckhoff catheter for peritoneal dialysis--an appraisal.
- Author
-
Rubin J, Adair CM, Raju S, and Bower JD
- Subjects
- Adolescent, Adult, Aged, Catheterization adverse effects, Child, Female, Humans, Infections etiology, Male, Middle Aged, Peritoneal Dialysis adverse effects, Peritonitis etiology, Prospective Studies, Catheterization instrumentation, Peritoneal Dialysis instrumentation
- Abstract
We prospectively evaluated early (within 40 days) catheter complications in all patients receiving a dialysis catheter between 1/8/80 and 1/8/81. 50% of patients achieved a functioning catheter at the first insertion and 24% required replacement of the catheter because of poor dialysate flow. Leaking from the catheter exit site occurred in 20%, infection at the exit site in 9% and peritonitis in 19% of patients. In patients who maintain a catheter over 40 days and undergo treatment by long-term peritoneal dialysis median catheter survival was 400 days with delayed cytheter failure primarily due to failure to resolve a clinical episode of peritonitis. Although the Tenckhoff catheter is readily inserted frequent complications occur.
- Published
- 1982
- Full Text
- View/download PDF