1. Biomechanical and functional variation in rat sciatic nerve following cuff electrode implantation.
- Author
-
Restaino, Stephen M., Abliz, Erkinay, Wachrathit, Kelliann, Krauthamer, Victor, and Shah, Sameer B.
- Subjects
NERVOUS system ,MULTIVARIATE analysis ,ELECTRODES ,NEUROLOGY ,PHYSIOLOGY - Abstract
Background Nerve cuff electrodes are commonly and successfully used for stimulating peripheral nerves. On the other hand, they occasionally induce functional and morphological changes following chronic implantation, for reasons not always clear. We hypothesize that restriction of nerve mobility due to cuff implantation may alter nerve conduction. Methods We quantified acute changes in nerve-muscle electrophysiology, using electromyography, and nerve kinematics in anesthetized Sprague-Dawley rat sciatic nerves during controlled hindlimb joint movement. We compared electrophysiological and biomechanical response in uncuffed nerves and those secured within a cuff electrode using analysis of variance (ANOVA) and regression analysis. Results Tethering following cuff implantation resulted in altered nerve strain and a complex biomechanical environment during joint movement. Coincident with biomechanical changes, electromyography revealed significantly increased variability in the response of conduction latency and amplitude in cuffed, but not free, nerves following joint movement. Conclusion Our findings emphasize the importance of the mechanical interface between peripheral nerves and their devices on neurophysiological performance. This work has implications for nerve device design, implantation, and prediction of long-term efficacy. [ABSTRACT FROM AUTHOR]
- Published
- 2014
- Full Text
- View/download PDF