1. Wafer level hermetic package and device testing of a SOI-MEMS switch for biomedical applications
- Author
-
Nicolaas F. de Rooij, Rogier Receveur, Vincent Larik, M. Zickar, and Cornel Marxer
- Subjects
Microelectromechanical systems ,Materials science ,business.industry ,Mechanical Engineering ,Electrical engineering ,Silicon on insulator ,Temperature cycling ,Hermetic seal ,Electronic, Optical and Magnetic Materials ,law.invention ,Reliability (semiconductor) ,Mechanics of Materials ,law ,Chip-scale package ,Shear strength ,Optoelectronics ,Wafer ,Electrical and Electronic Engineering ,business - Abstract
We have designed a wafer level chip scale package for a bi-stable SOI-MEMS dc switch using a silicon-glass hermetic seal with through the lid feedthroughs. Bonded at 365 °C, 230 V and 250 kg, they pass the fine/gross leak test after thermal cycling and mechanical shock/vibration according to MIL-STD-833, fulfilling the requirements for biomedical applications. The measured shear strength is 114 ± 26 N in correspondence with the theoretically expected 100 N. Ruthenium microcontacts are a factor of 100 more robust than gold microcontacts, being stable over 106 cycles measured in a N2 atmosphere inside the package presented here. Future work will include a more extensive bond quality assessment and continued microcontact reliability measurements.
- Published
- 2006
- Full Text
- View/download PDF