1. A prospective comparison of the platelet sequestration ability of three autotransfusion devices.
- Author
-
Scholz M, Serrick C, Noel D, Singh O, and Melo A
- Subjects
- Adult, Blood Transfusion, Autologous methods, Hematocrit, Humans, Platelet Transfusion methods, Prospective Studies, Thoracic Surgery instrumentation, Treatment Outcome, Blood Platelets, Blood Transfusion, Autologous instrumentation, Platelet Count, Platelet Transfusion instrumentation, Thoracic Surgery methods
- Abstract
Although current autotransfusion devices have platelet sequestration capabilities, each has a unique technology to achieve the final platelet product. The purpose of this study was to assess the quality and quantity of platelets sequestered by three different autotransfusion devices. The three commercially available autotransfusion devices evaluated were Fresenius C.A.T.S (closed spiral chamber), Cobe BRAT 2 (Baylor bowl), and Haemonetic Cell Saver 5 (Latham bowl). Platelet sequestration was preformed in the automatic mode following the manufacturer's recommended sequestration protocols. The total number of platelets sequestered, percent recovery, and percent platelet function were assessed. Each device behaved similarly. There was a 2- to 3-fold increase in platelet count compared with baseline with only approximately 50-60% recovery, whereas there was approximately a 10% decrease in platelet function after processing compared with baseline. No statistical difference was noted in platelet function between the respective machines. However, there was a significant loss of platelet function observed with the actual process regardless of autotransfusion device used.
- Published
- 2005