1. Validation of a simplified procedure for convenient and rapid quantification of reduced and oxidized glutathione in human plasma by liquid chromatography tandem mass spectrometry analysis.
- Author
-
Enomoto AC, Schneider E, McKinnon T, Goldfine H, and Levy MA
- Subjects
- Glutathione blood, Humans, Limit of Detection, Linear Models, Reproducibility of Results, Chromatography, Liquid methods, Glutathione Disulfide blood, Tandem Mass Spectrometry methods
- Abstract
Endogenous glutathione (GSH) and glutathione disulfide (GSSG) status is highly sensitive to oxidative conditions and have broad application as a surrogate indicator of redox status in vivo. Established methods for GSH and GSSG quantification in whole blood display limited utility in human plasma, where GSH and GSSG levels are ~3-4 orders of magnitude below those observed in whole blood. This study presents simplified sample processing and analytical LC-MS/MS approaches exhibiting the sensitivity and accuracy required to measure GSH and GSSG concentrations in human plasma samples, which after 5-fold dilution to suppress matrix interferences range from 200 to 500 nm (GSH) and 5-30 nm (GSSG). The utility of the methods reported herein is demonstrated by assay performance and validation parameters which indicate good sensitivity [lower limits of quantitation of 4.99 nm (GSH) and 3.65 nm (GSSG), and high assay precision (intra-assay CVs 3.6 and 1.9%, and inter-assay CVs of 7.0 and 2.8% for GSH and GSSG, respectively). These methods also exhibited exceptional recovery of analyte-spiked plasma samples (98.0 ± 7.64% for GSH and 98.5 ± 12.7% for GSSG). Good sample stability at -80°C was evident for GSH for up to 55 weeks and GSSG for up to 46 weeks, with average CVs <15 and <10%, respectively., (© 2020 The Authors. International Journal for Numerical Methods in Engineering published by John Wiley & Sons Ltd.)
- Published
- 2020
- Full Text
- View/download PDF