1. Temperature manifold for a stopped-flow machine to allow measurements from −10 to +40 °C.
- Author
-
Walklate, Jonathan and Geeves, Michael A.
- Subjects
- *
FLOW injection analysis , *MYOSIN , *LABORATORY rabbits , *PROTEIN analysis , *ENZYME activation - Abstract
Conducting enzymatic stopped-flow experiments at temperatures far removed from ambient can be very problematic because extremes in temperature (<10 °C or >30 °C) can damage the machine or the enzyme. We have devised a simple manifold that can be attached to most commercial stopped-flow systems that is independently heated or cooled separate from the main stopped-flow system. Careful calibration of the flow circuit allows the sample to be heated or cooled to the measurement temperature (−8 to +40 °C) 1 to 2 s before mixing in the reaction chamber. This approach allows measurements at temperatures where the stopped flow or the protein is normally unstable. To validate the manifold, we investigated the well-defined ATP-induced dissociation of rabbit muscle myosin subfragment 1 (S1) from its complex with pyrene-labeled actin. This process has both temperature-dependent and -independent components. Use of ethylene glycol allowed us to measure the reaction below 0 °C and up to 42 °C, and as expected the second-order rate constant ( K 1 k +2 ) and the maximum rate of dissociation ( k +2 ) both increased with temperature, whereas 1/ K 1 is unaffected by the change in temperature. [ABSTRACT FROM AUTHOR]