1. A FPGA-Based Broadband EIT System for Complex Bioimpedance Measurements -- Design and Performance Estimation
- Author
-
Kusche, Roman, Malhotra, Ankit, Ryschka, Martin, Ardelt, Gunther, Klimach, Paula, and Kaufmann, Steffen
- Subjects
Physics - Medical Physics - Abstract
Electrical impedance tomography (EIT) is an imaging method that is able to estimate the electrical conductivity distribution of living tissue. This work presents a field programmable gate array (FPGA)-based multi-frequency EIT system for complex, time-resolved bioimpedance measurements. The system has the capability to work with measurement setups with up to 16 current electrodes and 16 voltage electrodes. The excitation current has a range of about 10 $\mu$A to 5 mA, whereas the sinusoidal signal used for excitation can have a frequency of up to 500 kHz. Additionally, the usage of a chirp or rectangular signal excitation is possible. Furthermore, the described system has a sample rate of up to 3480 impedance spectra per second (ISPS). The performance of the EIT system is demonstrated with a resistor-based phantom and tank phantoms. Additionally, first measurements taken from the human thorax during a breathing cycle are presented.
- Published
- 2020
- Full Text
- View/download PDF