1. Accuracy of T2 magnetic resonance assays as point-of-care methods in the intensive care unit.
- Author
-
Vrettou, C.S., Douka, E., Perivolioti, E.P., Vassiliou, A.G., Sarri, A., Giannopoulou, V., Trigkidis, K.K., Jahaj, E., Dimopoulou, I., and Kotanidou, A.
- Abstract
Novel molecular diagnostic methods are being evaluated in order to expedite pathogen identification in patients with bacteraemia. To evaluate the feasibility and diagnostic accuracy of the T2 magnetic resonance (T2MR) assays – T2 Bacteria (T2B) and T2 Resistance (T2R) – as point-of-care tests in the intensive care unit compared with blood-culture-based tests. Prospective cross-sectional study of consecutive patients with suspected bacteraemia. Diagnostic accuracy was evaluated using blood culture as the reference method. In total, 208 cases were included in the study. The mean time from sampling to report was lower for the T2MR assays compared with blood-culture-based methods (P <0.001). The rate of invalid reports was 6.73% for the T2B assay and 9.9% for the T2R assay. For the T2B assay, overall positive percentage agreement (PPA) was 84.6% [95% confidence interval (CI) 71.9–93.1%], negative percentage agreement (NPA) was 64.3% (95% CI 55.4–72.6%), positive predictive value (PPV) was 48.9% (95% CI 42.5–55.3%) and negative predictive value (NPV) was 91.2% (95% CI 84.4–95.2%). Cohen's kappa coefficient was 0.402. For the T2R assay, overall PPA was 80% (95% CI 51.9–95.7%), NPA was 69.2% (95% CI 54.9–81.3%), PPV was 42.9% (95% CI 31.7–54.8%) and NPV was 92.3% (95% CI 81.1–97.1%). Cohen's kappa coefficient was 0.376. T2MR assays have high NPV for rapid exclusion of bacteraemia, and could potentially assist with antimicrobial stewardship when applied as point-of-care diagnostic tests in the intensive care unit. [ABSTRACT FROM AUTHOR]
- Published
- 2023
- Full Text
- View/download PDF