1. Development and Validation of RP-HPLC Method for the Determination of Methamphetamine and Propranolol in Tablet Dosage Form
- Author
-
Shabir, G. A.
- Subjects
test method validation ,reversed-phase high-performance liquid chromatography ,tablets ,propranolol ,Research Paper ,Methamphetamine - Abstract
A new isocratic reversed-phase HPLC method with diode-array UV detection was developed and validated for the determination of methamphetamine and propranolol in tablet dosage forms. Chromatography was carried out on an XTerra RP18 (150×4.6 mm, 5 μm) column using 50 mM pyrrolidine (pH 11.5) - acetonitrile (50:50, v/v) as mobile phase at a flow rate of 1 ml/min. Spectrophotometric detection was performed at a wavelength of 214 nm. The linearity was established over the concentration range of 0.075-0.60 mg/ml for both drugs. The correlation coefficients (r(2)) were ≥0.9998 in each case. The relative standard deviation values for intermediate precision studies were1%. Statistical analysis of the data showed that the method was precise, accurate, reproducible and selective for the analysis of methamphetamine and propranolol drugs. The method was successfully employed for the determination of propranolol and methamphetamine in commercially available tablet dosage form.
- Published
- 2011